About Test Equipment
Q55: Is there any provision on the diameter of the wheel attachment surface in the attachment jig of the test equipment?
Q56: To what extent do calibration and inspection need to be performed on dead weight dynamic cornering fatigue test equipment?
Q57: Do accuracy check reports on the testing equipment need to be submitted every year?
Q58: Is it permissible to perform accuracy checks of testing equipment using our own master gauge block?
Q59: Testing equipment with a certified ID was given to us. Is it possible to undergo documentary judgment using this testing equipment?

Q55: Is there any provision on the diameter of the wheel attachment surface in the attachment jig of the test equipment?
A55: There is no specific provision. However, generally jigs are used with attachment surfaces of a larger diameter than that of the disc attachment surface of the wheel to be tested. Jigs for attachment should not interfere with the wheel.


Q56: To what extent do calibration and inspection need to be performed on dead weight dynamic cornering fatigue test equipment?
A56: Calibration and inspection need to be performed on the weight and arm length.


Q57: Do accuracy check reports on the testing equipment need to be submitted every year?
A57: Yes. Since maintenance of accuracy and ensuring performance of the certified testing site and equipment through maintenance and inspection is indispensable, inspections and reporting are assumed to be carried out every year.


Q58: Is it permissible to perform accuracy checks of testing equipment using our own master gauge block?
A58: It is permissible if your master gauge block is traceable by the National Master Gauge Block.


Q59: Testing equipment with a certified ID was given to us. Is it possible to undergo documentary judgment using this testing equipment?
A59: It is necessary to check how the equipment is placed and managed at the company the equipment was transferred from. Therefore equipment inspection (on-site inspection) will be required again. Application for approval of testing equipment should be made newly, and after completion of approval and registration based on the equipment, judgment will be completed and testing will be conducted, and then after compliance with the standards is verified, please apply for registration via documentary judgment.